EDWARDS LIFESCIENCES SAPIEN 3 TRANSCATHETER HEART VALVE; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9600TFX29
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 11/17/2022
Event Type  Injury  
Manufacturer Narrative
Per the instructions for use (ifu), valve thrombosis is a potential risk associated with the use of the transcatheter heart valves (thv).Thrombosis is a rare and well-recognized complication associated with the use of bioprosthetic heart valves.Valve thrombosis is the formation of significant blood clots forming on the valve.Potential patient risk factors such as atrial fibrillation, systemic disease (e.G., systemic lupus erythematosus, inflammation and damage to various body tissues, including joints, skin, kidneys, heart, lungs, blood vessels, and brain), the valvular disease itself, and reduced cardiac ejection fraction can contribute to increased risk of thrombus/thrombosis.Sub-optimal anticoagulation during the procedure and underlying patient conditions can also result in increased thrombogenicity.These patients are anticoagulated for the procedure and interventional best practices mandate meticulous preparation, aspiration, and flushing of the devices to prevent and/or remove clot(s).Short-term anticoagulation therapy may also be necessary after valve repair, anticoagulation is prescribed per institutional guidelines.Intra-procedural intra-cardiac thrombus and post-procedure or late valve thrombosis are complex processes triggered by the interaction between the host and the device which are highly variable among patients.It is the natural tendency of the body to form a clot on foreign objects in the vascular space and a definitive root cause cannot always be confirmed.The device training manuals instruct the operator to consider all procedural and anatomical factors.Physicians are extensively trained by edwards before they are qualified to use the device system.Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.Operators are also instructed to use fluoroscopy in conjunction with echocardiography for optimal visualization during positioning and deployment and cautions include the maintenance of the patient's anticoagulation status (act at >250 seconds) during the procedure.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results are inconclusive as patient and procedural factors were not provided; however, the event could be related to the mechanisms described above.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.Valve remains implanted.
 
Event Description
During tmvr procedure using a 29mm sapien 3 (s3) case by transfemoral approach in mitral position inside a pre-existing 29mm perimount valve (viv procedure).Initially, it was very difficult to pass the surgical perimount valve with the s3.After successful implantation, the patient's pressure dropped down and reanimation was provided.Pericardial tamponade was recognized by echo and treated with a "punction".The patient was connected to ecmo and converted to cardiac surgery.A cut from the guidewire was found in the left apex and it was sutured.After successful closure, the echo showed a non-working sapien 3 valve.A thrombus acutely disabled the function of the 29mm s3 valve.A claret was introduced, again ts ballooning (2 x) with a 25 mm edwards balloon catheter of the s3 in the mitral position to mobilize the leaflets.The thrombus was still recognized with a semi-functioning s3.The procedure was finished and, again, the pressure dropped down and pericardial tamponade was found by echo.It was decided, together with the family of the patient, to stop active therapy regarding the advanced decision of the patient.The patient is still alive in the icu.
 
Manufacturer Narrative
This is one of two manufacturer reports being submitted for this case.Please reference the related manufacturer report no: 2015691-2022-09915.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAPIEN 3 TRANSCATHETER HEART VALVE
Type of Device
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key15956318
MDR Text Key305243179
Report Number2015691-2022-09913
Device Sequence Number1
Product Code NPU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/06/2023
Device Model Number9600TFX29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2022
Initial Date FDA Received12/11/2022
Supplement Dates Manufacturer Received12/12/2022
Supplement Dates FDA Received12/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-