| Model Number |
PRT-29 |
| Medical Device Problem Code |
Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Code |
Arrhythmia (1721)
|
| Date of Event |
12/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
Crd_669 ¿ portico i, patient site id: (b)(4).It was reported that on (b)(6) 2016, a 29mm portico valve was implanted.There was good results post-implant with small para and transvalvular insufficiency.On (b)(6) 2021, echocardiogram was performed and revealed extrasystoles and only mild regurgitation.The patient was administered betablocker therapy.The patient is reported to be stable.It is believed that the extrasystoles are due to patient condition and not device performance.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of extrasystoles and mild regurgitation was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the patient had baseline conduction abnormalities which included hypertension,hypertrophic cardiomyopathy and left bundle branch block.The field also indicated that it was believed that the reported event was caused by patient condition.Based on the information received the cause of the reported incident could not be conclusively determined but could have been as a result of patient conditions.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Na.
|
| |
|
Search Alerts/Recalls
|