• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown as information was not provided, the best estimate date is between 28-mar-2022 and 14-nov-2022.The device was not returned for analysis.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain additional information regarding the event; however, the information was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the zxt300 14.0 diopter intraocular lens (iol) was explanted from the patient's ocular dexter (right eye) in a secondary surgical procedure due to poor vision.A different lens, model diu300 of 14.0 diopter was used as the replacement.There were no complications, such as capsular tear, incision enlargement, suture(s) and/or vitrectomy reported.Patient outcome post-lens exchange was good, and no serious injury reported.No further information is available.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 28-dec-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received cut into three pieces.The lens was cleaned and, no issues that could contribute to or cause the complaint issues could be observed.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issues were not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15956434
Report Number3012236936-2022-02953
Device Sequence Number5526250
Product Code POE
UDI-Device Identifier05050474606265
UDI-Public(01)05050474606265(17)220522
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/22/2022
Device Model NumberZXT300
Device Catalogue NumberZXT300U140
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date12/11/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-