Model Number EV07362CD
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.Expiration date: 07/2024.Device not returned.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the stent allegedly dislodged.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the stent allegedly dislodged.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One image was reviewed.One fluoroscopic image is reviewed.This shows a stent within a tibial artery.The lack of imaging quality makes it difficult to determine if this is appropriately expanded within the artery, although no major defects are appreciated.It appears that the interventionalist was able to post dilate the stent despite maldevelopment.Further media is needed to confirm the exact event that led to this report.Therefore, the investigation for the reported stent dislodgment remains inconclusive as no objective evidence could be found during the image review.A definitive root cause for the reported stent dislodgment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALEO BALLOON EXPANDABLE BILIARY STENT
Common Device Name
BALLOON EXPANDABLE STENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15957432
Report Number2020394-2022-00975
Device Sequence Number20110481
Product Code FGE
UDI-Device Identifier00801741000768
UDI-Public(01)00801741000768
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K052132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEV07362CD
Device Catalogue NumberEV07362CD
Device Lot NumberGFGT1836
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight97 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-