| Model Number |
EUP3015X |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An attempt was made to use one euphora device to treat a non tortuous, non calcified lesion.No damage was noted to the packaging, and the device was removed from the hoop with no issues noted.The device was inspected with no issues.Negative prep was performed with no issues noted.The lesion was not pre-dilated.The device did not pass through a previously-deployed stent.Resistance was not encountered when advancing the device and excessive force was not used during delivery.It was reported that the catheter detached, cracked or fractured during removal of the device.The device was not kinked and re straightened during use.It was stated that the device does not remain in the patient, that it was snared out.It was stated that the balloon was used as a trapping balloon and was never placed in an artery.The patient is reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the device was received for analysis.The device was not detached at transition shaft from the hypotube on return, as the account had placed the detached transition shaft back onto the hypotube.The detachment of the transition shaft from the hypotube was evident approx.41.5cm distal to the strain relief.The detachment of the transition shaft may have been due to the hypotube bond not being formed correctly.The transition shaft material was kinked and flattened along its length.Marks were visible on the shaft material at the detachment site.Kinks were evident to the distal shaft.Blood was visible in the inflation lumen.Blood was visible in the balloon.The balloon folds were expanded on return.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: it was reported that the mid shaft of the catheter detached, cracked or fractured during removal of the device from the guide catheter.The balloon was being used as a trapper balloon.It was advanced through the guide not over a wire, placed and inflated at the tip of the guide to hold the wire in place as the microcatheter was removed from the artery.The euphora device was never placed in an artery.Image review: the image provided shows a detachment.The core wire is visible indicating that the detachment occurred on the transition shaft.Blood is visible in the inflation lumen and balloon.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|