• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN046313
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Headache (1880)
Date of Event 11/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
It was reported that the epidural set syringe not generally vacuum to identify epidural space.Plungers loose and falling off.The device was removed entirely and replaced with a different brand (b braun) successfully.There was no serious injury, complications, or delay caused to the patient.Only exacerbation of headache due to multiple attempts.
 
Event or Problem Description
It was reported that the epidural set syringe not generally vacuum to identify epidural space.Plungers loose and falling off.The device was removed entirely and replaced with a different brand (b braun) successfully.There was no serious injury, complications, or delay caused to the patient.Only exacerbation of headache due to multiple attempts.
 
Additional Manufacturer Narrative
(b)(4).Visual inspection could not be performed as no sample was returned for analysis.The customer did provide two photos that appear to show a sealed epidural catheter kit.A dimensional or functional inspection could not be performed as no sample was returned for analysis.Additional testing could not be performed as no sample was returned for analysis.A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part #kz-05501-002 as a part of this complaint investigation.There have been no material changes for this part during the last two years that could have led to this complaint.A corrective action is not required at this time as the root cause for this complaint investigation could not be determined based upon the information provided and without the actual sample.Complaint verification testing could not be performed as no sample was provided for analysis.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key15958097
Report Number3006425876-2022-01128
Device Sequence Number14690319
Product Code BSO
UDI-Device Identifier00801902062512
UDI-Public00801902062512
Combination Product (Y/N)N
Initial Reporter CountryZI
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberIPN046313
Device Catalogue NumberJC-05400-DCS
Device Lot Number71F21H1401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient Age36 YR
Patient SexUnknown
-
-