| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2022 - 00493, 3002806535 - 2022 - 00494.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised from an oxford uni to a full knee due to patella pain.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: oxf uni tib tray sz b rm pma item#154721 lot#016760.Oxf twin-peg cmntd fem sm pma item#161468 lot#441070.This follow-up report is being submitted to relay additional and corrected information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Devices are used for treatment.Radiographs were provided but as they appear to be xeroxed or faxed images and are grainy in quality.A hcp assessment will not enhance the investigation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|