• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS HEMAGARD KNITTED BIFURCATED; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS HEMAGARD KNITTED BIFURCATED; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HGK1407
Medical Device Problem Code Material Too Soft/Flexible (4007)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The actual device involved in the adverse event remained implanted.
 
Event or Problem Description
It was reported to intervascular that the implanted graft was a substitution for "#(b)(4)".They used one [graft] on (b)(6) 2022 and the week before, and stated "the graft was very soft and flimsy and kept kinking in the tunnel.Leading to 3 repeated attempts and case bleeding.It is not a safe and acceptable alternative".Complaint #(b)(4).
 
Additional Manufacturer Narrative
Corrected data : block b1"product problem" was updated to "product problem/ adverse event" due to the lack of information, the worst case scenario was considered.Block b2 was completed as required.(4111) several attempts to contact the customer was performed by the medical department.No response could be obtained.(4112/213) the case was reviewed by the corporate medical department whose assessment is below: "the complaint received states that a hemagard 14x7 bifurcated graft was implanted in a patient.It appears that the hemagard graft is a substitute for their typically used graft, however the brand of graft is unknown at this time.The customer states that the hemagard graft is soft, flimsy and kinks easily, causing bleeding after repeated attempts to implant.No other information is known at this time.Retention product from the same lot, coated on the same day was visually inspected and underwent water permeability testing and revealed that the product is within specifications.A request for additional details of the procedure has been sent back to the customer, and these comments and summary will be updated when they are received.Attempts to request further information on this complaint were made on dec 14, 2022; dec 15, 2022; dec 22, 2022; and jan 5, 2023.There was no response.Due to the lack of information available we were unable to arrive at a conclusion.Should additional information be obtained in the future, this medical assessment will be updated to reflect those findings." (4315) no conclusion can be drawn on the exact origin of the adverse event since the product remained implanted.However, the conducted investigation and testing performed suggest that the product was not defective at the time of manufacturing.
 
Event or Problem Description
Complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD KNITTED BIFURCATED
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR  13705
Manufacturer (Section G)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR   13705
Manufacturer Contact
clemence vaneenoge
zone industrielle athelia i
la ciotat 13705
FR   13705
442084646
MDR Report Key15958410
Report Number1640201-2022-00037
Device Sequence Number17420038
Product Code DSY
UDI-Device Identifier00384401014386
UDI-Public(01)00384401014386
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K964625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHGK1407
Device Catalogue NumberHGK1407
Device Lot Number22A20
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer01/14/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-