• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 364957
Device Problem Physical Resistance/Sticking (4012)
Patient Problem Needle Stick/Puncture (2462)
Event Date 11/16/2022
Event Type  Injury  
Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event Description
It was reported when using the bd vacutainer® urine complete cup kit there was a needle stick injury - post use.The following information was provided by the initial reporter.The customer stated: "tech poke his finger with needle (transfer device).Tech decline any medical testing and/or intervention".
 
Event Description
It was reported when using the bd vacutainer® urine complete cup kit there was a needle stick injury - post use.The following information was provided by the initial reporter.The customer stated: "tech poke his finger with needle (transfer device).Tech decline any medical testing and/or intervention.".
 
Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to needle exposure as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode needle exposure.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® URINE COMPLETE CUP KIT
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15958759
MDR Text Key305250119
Report Number1917413-2022-00776
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K024240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number364957
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-