MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 6IN AML LRG STATURE 12.0 MM; AML IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 6IN AML LRG STATURE 12.0 MM; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 1554-02-121
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994)
Date of Event 11/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was presented with mid thigh pain.X-rays showed a fracture in the stem roughy where the flange begins.Patient says no trauma occurred, was walking around on vacation.After re-examining x-rays from 2018, it appears there may have been the beginning of a fracture.At the time, patient presented with bursitis symptoms.Current x-rays show remodeling lateral to the stem fracture.No sign of infection.No surgical delay was reported.Doi: (b)(6) 2008.Dor: (b)(6) 2022.Affected side: left.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot
=
> a manufacturing record evaluation (mre), was not possible because the provided lot code was not valid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6IN AML LRG STATURE 12.0 MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15958781
Report Number1818910-2022-25033
Device Sequence Number12156194
Product Code LPH
UDI-Device Identifier10603295058359
UDI-Public10603295058359
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K061833
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1554-02-121
Device Catalogue Number155402121
Device Lot NumberD155402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6IN AML LRG STATURE 12.0 MM; ARTICULEZE M HEAD 36MM +8.5; PINNACLE MTL INS NEUT36IDX52OD
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-