MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS SO 12.0; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS SO 12.0; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-106120
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while inserting the stem, the doctor noted that it looked like the grit blast was coming off the stem.There were no consequences or impact to the patient.Attempts have been made and no further information is available.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: cat#: 31-323220/ 3.2mmx20mm rnglc+ acet drl bit/lot#: 546040.Cat#: 110010244/ g7 osseoti 3 hole shell 52mm e/ lot#: 65459659.Cat#: 30103605/ g7 vit e neutral lnr 36mm e/lot #: 65560835.Cat#: 00-6250-065-20/ bone screw 6.5x20 selftap/ lot#: j7238089.Cat#: 00-6250-065-20/ bone screw 6.5x20 selftap/ lot#: j7348330.Cat#: 51-106120/ tprlc 133 mp type1 pps so 12.0/lot#: 7041275.Cat#: 650-0661/ delta ceramic fem hd 36/0mm/lot#: 3119037.Cat#: 98-0001-000-37/ pr prm st/oss lh cp/ve ln/cer/ins.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: h2; h3; h6 component code: mechanical (g04) ¿ stem no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.The device history records were reviewed for deviations and/or anomalies with the following related anomalies/deviations identified: the auto blaster was down so they had to use a manual blaster, however, the device inspection showed the device was conforming per spec.Therefore, it is unlikely that the deviation contributed to the reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS SO 12.0
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15958800
Report Number0001825034-2022-02771
Device Sequence Number5526318
Product Code KWA
UDI-Device Identifier00880304514317
UDI-Public(01)00880304514317(17)310611(10)7041275
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-106120
Device Lot Number7041275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexMale
-
-