Model Number 10590000
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to siemens that an adverse event and malfunction occurred while operating the somatom definition edge ct system.On (b)(6)2022, a scan was performed on a pediatric male patient.The customer reported that the user interface of the scanner was not responding after the topogram (low dose overview scan).As a result, the patient was rescanned on another ct system.Beside the additional x-ray dose, no other injury was reported.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a software error.The application freezes because the csapixelframe object of the topogram image is deleted while the image is released from the image processing toolkit (ipt) server and the image is still being processed in the background.After reboot of the system, the functionality recovers.This identified software issue has already been resolved with the service pack vb20_sp6 and released for distribution with ct014/22/s update vb20_sp6 on (b)(6)2022.This system is adressed for the update.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION EDGE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1 -or-
rittigfeld 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key15959004
Report Number3004977335-2022-52989
Device Sequence Number5573759
Product Code JAK
UDI-Device Identifier04056869006949
UDI-Public04056869006949
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other
Remedial Action Inspection,Recall
Type of Report Initial
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10590000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Initial Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age7 YR
Patient SexMale
Patient Weight20 KG
-
-