• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 40MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 40MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 623-00-40E
Medical Device Problem Codes Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
It was reported by the patient's attorney as a result of a legal claim that allegedly the patient was implanted with an v40 cocr lfit femoral head on his left hip on (b)(6) 2009.It is further alleged that the patient experienced pain.X-rays allegedly showed "broken, worn out trunnion and the acetabular socket appeared to be in good condition." during revision surgery on (b)(6) 2020, "the worn out trunnion was completely corroded and wearing on a superior portion of the socket.The plastic liner was broken in this area as well.".
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation of the metal head from the accolade stem and crack/fracture of the accolade stem involving a trident liner was reported.Conclusion: based on the information provided there is no indication or allegation that the device reported in this investigation contributed to the event and is therefore considered concomitant.The patient was revised due to disassociation of the head from the stem and fracture of the stem trunnion.There were no reported allegations against the trident liner and it was only reported to be revised due to damage from articulation against the disassociated stem trunnion.A full investigation is completed on the accolade stem and metal head.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported by the patient's attorney as a result of a legal claim that allegedly the patient was implanted with an v40 cocr lfit femoral head on his left hip on (b)(6) 2009.It is further alleged that the patient experienced pain.X-rays allegedly showed "broken, worn out trunnion and the acetabular socket appeared to be in good condition." during revision surgery on (b)(6) 2020, "the worn out trunnion was completely corroded and wearing on a superior portion of the socket.The plastic liner was broken in this area as well.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 40MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15959068
Report Number0002249697-2022-01801
Device Sequence Number5526327
Product Code LPH
UDI-Device Identifier07613327025323
UDI-Public07613327025323
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K062419
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2014
Device Model Number623-00-40E
Device Catalogue Number623-00-40E
Device Lot NumberMHLT0V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-