There was an allegation of questionable gen.2 ise indirect for na results for 3 patient samples on a cobas integra 400 plus module.The customer was prompted to rerun patient samples due to qc and calibration failures from a new na electrode.For patient 1, the initial na result was 124 mmol/l.The repeat result was 131 mmol/l.The initial result was reported outside of the laboratory.The repeat result was deemed correct.On (b)(6) 2022, the initial na results for patient 2 were 129.3 mmol/l and 129.3 mmol/l.On (b)(6) 2022, the repeat results were 134.4 mmol/l, 134.6 mmol/l, and 133.9 mmol/l.On (b)(6) 2022, the initial na results for patient 3 were 130.3 mmol/l, 128.0 mmol/l, and 129.0 mmol/l.On (b)(6) 2022, the repeat result was 134.6 mmol/l.It is unknown if the initial results for patients 2 and 3 were reported outside of the laboratory.The na electrode lot number and expiration date were requested but not provided.
|