| Model Number |
498106 |
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
12/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sample is unavailable for evaluation.Without such evident to review, the complaint cannot be confirmed.If additional information is provided in the future, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The connection point between the intra-arterial catheter and the stabilizing devise was broken and the catheter was retained in the artery.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.81 sealed samples were returned for review.Per the product experience report, there was no sample available for review and no defective product was returned with the sealed samples.Five random sealed samples were opened by the analyst and a pull test was conducted with no failures per op-qa-007 which allows for a destructive pull test of 1.75 lbs.Since the reported issue could not be duplicated per an examination of the returned product, no further evaluation was conducted.There have been no other complaints regarding this issue with this lot number.Since the reported issue could not be duplicated per an examination of the returned product, no further evaluation was conducted and no corrective action was taken.Argon will continue to monitor for issues of this nature in the future.
|
| |
|
Event or Problem Description
|
|
The connection point between the intra-arterial catheter and the stabilizing devise was broken and the catheter was retained in the artery.
|
| |
|