| Model Number |
100/189/085 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when the customer put air in the cuff during the pre-use check, leakage of air from it was observed.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device serial number/lot number is unknown, no product information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: two photos were included in agile attachments for evaluation; photos showed the complainant sample.One sample was returned for investigation.The returned sample was received in used condition without original packaging.The sample was visually inspected under normal conditions of illumination to detect any condition that could cause the failure mode reported.A leakage was detected between the check valve and pilot balloon.No other analysis was performed.Based on the analysis conducted in the sample provided, a leaking failure mode was confirmed.The failure condition was possibly due to assembly process not followed correctly (not enough solvent at joint).The problem source was traced to manufacturing.For corrective action an awareness/customer complaint notification was performed to production personnel to explain the importance of adherence to the manufacturing procedure.No lot number was provided; therefore, device history record review could not be completed.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).
|
| |
|
Search Alerts/Recalls
|