• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number ALLURA XPER FD20/15 OR TABLE
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported to philips that the system produced an error of ¿generator busy¿ during start up.There was no reported patient or user harm.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed the error generator busy.Review of the system log file showed com xsc: tube current out of range and x-ray generator: phase fault.Upon further inspection, fse performed a tube adaptation.After performing the tube adaptation, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key15961634
Report Number3003768277-2022-01475
Device Sequence Number17378182
Product Code OWB
UDI-Device Identifier00884838059122
UDI-Public00884838059122
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141979
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberALLURA XPER FD20/15 OR TABLE
Device Catalogue Number722059
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer12/21/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date12/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-