| Model Number |
ALLURA XPER FD20/15 OR TABLE |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It has been reported to philips that the system produced an error of ¿generator busy¿ during start up.There was no reported patient or user harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed the error generator busy.Review of the system log file showed com xsc: tube current out of range and x-ray generator: phase fault.Upon further inspection, fse performed a tube adaptation.After performing the tube adaptation, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
|
| |
|