• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MX-15-T
Medical Device Problem Codes Deflation Problem (1149); Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Failure of Implant (1924); Seroma (2069)
Date of Event 11/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported unknown side "noticed a hole in the seam".Device was explanted and replaced with a backup.
 
Event or Problem Description
Healthcare professional reported left side "noticed a hole in the seam".Healthcare professional additionally reported left side "some serous fluid within the tissue expander.".
 
Additional Manufacturer Narrative
In response to fda report number: (b)(4).Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed opening device in the stable base-shell bond edge side assessed as an unidentified (tear) opening.Shell thickness within specification).Use error: unable to confirm as it is a medical event not related to the device.Additional observations: white particles observed in the surface of the shell.No further actions are required as the device was implanted.Additional, changed, and/or corrected data: b5, d6a, d9, e4, g2, h3, h6.The event of "seroma" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed opening device in the stable base-shell bond edge side assessed as an unidentified (tear) opening.Shell thickness within specification).Use error: unable to observe as it is related to the time implanted and is not related to the device.Seroma: unable to observe as it is a medical event not related to the device.Additional observations: white particles observed in the surface of the shell.No further actions are required as the device was implanted.
 
Event or Problem Description
Healthcare professional reported left side "noticed a hole in the seam".Later, healthcare professional reported that there was no mention of seroma on operating procedure report; therefore, seroma was removed as an event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MX-15-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15962137
Report Number9617229-2022-21850
Device Sequence Number20141074
Product Code LCJ
UDI-Device Identifier10888628044210
UDI-Public10888628044210
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/07/2025
Device Model Number133S-MX-15-T
Device Catalogue Number133S-MX-15-T
Device Lot Number3419825
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer12/20/2022
01/17/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date01/13/2023
02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexUnknown
Patient Weight86 KG
-
-