| Model Number |
2218-87-356 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hip revision: the surgeon was trialing with the ab 56/36mm trial into the gription acetabular cup (relocated and took the leg through a rom), and when they removed the trial from the acetabulum they noticed it had come apart on the mayo table.Saline was used in the patient with suction to clean any debris.Surgeon was happy.Was surgery delayed due to the reported event? no.Was procedure successfully completed? yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: the device associated with this report was not returned for analysis.Review of the photographic evidence confirmed the reported allegation.The snap ring was found broken, no other defects were observed on the device.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed as no valid finished good lot number was provided for this device.
|
| |
|
Search Alerts/Recalls
|