| Model Number |
72201092 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during set up or inspection, the footswitch was getting hot.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case-(b)(4).The reported device was received for evaluation.A visual inspection was performed and no deficiencies were observed.A functional evaluation was performed on the returned device and the reported malfunction was not observed.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|