MAUDE Adverse Event Report: STRYKER GMBH DISTAL LATERAL FEMUR PLATE AXSOS 3 FOR LEFT FEMUR 10 HOLE / L238MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DISTAL LATERAL FEMUR PLATE AXSOS 3 FOR LEFT FEMUR 10 HOLE / L238MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 627610S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "axsos plate broken 5 weeks after medical procedure.Follow-up surgery with t2 scn planned.".
 
Event or Problem Description
As reported: "axsos plate broken 5 weeks after medical procedure.Follow-up surgery with t2 scn planned.".
 
Additional Manufacturer Narrative
Correction: please refer to d9/h3 - the device was returned for investigation.H6 (results & conclusion codes).The reported event could be confirmed based on the investigation performed on the device.The device inspection revealed the following: the identification of the returned plate was confirmed based on the catalog # and the lot # marked.The scratches observed on the implant surface occurred most probably during the explanation.The breakage occurred in the distal part of the implant, both parts were returned.The microscopic inspection of the fracture face has shown that the typical characteristics of a fatigue fracture can be identified at the fracture face on the right side of the implant.The origin of the crack is at the notch intended for the assembly of the targeting arm.There is a fatigue zone with a smooth surface and progression lines over al-most the whole surface of the fracture and finally there is in instantaneous zone, where the final fracture occurred.On the fracture face on the center and left side, the shiny surface indicates that the fracture faces did rub against each other for a longer period.The rest of the fracture face has the typical view of a fatigue fracture with a uniform fine-grained texture.The presence of the notch leads to a material lessening, thus a creating weak spot in the plate.The fatigue fracture observed is commonly a result of bone delayed union/non union.However, as x-rays were not received for investigation, it was not possible to make any statement on the bone condition of the patient in this case.There was no information on the patient's health condition and post-operative behavior.Dimensional inspection confirmed that the device is within specification.More information such as x-rays and patient conditions are needed in order to determine the exact root cause of the event.However, based on the length of the implantation of the device, and the fracture mode, this event was most probably caused by bone delayed union/ malunion, which led to excessive loads being applied to the implant, causing its breakage.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Additional Manufacturer Narrative
Correction - please refer to d9/h3 - the device was not returned for investigation.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition is unknown.
 
Event or Problem Description
As reported: "axsos plate broken 5 weeks after medical procedure.Follow-up surgery with t2 scn planned.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL LATERAL FEMUR PLATE AXSOS 3 FOR LEFT FEMUR 10 HOLE / L238MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15963623
Report Number0008031020-2022-00670
Device Sequence Number5549766
Product Code HRS
UDI-Device Identifier04546540685681
UDI-Public04546540685681
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K200398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number627610S
Device Catalogue Number627610S
Device Lot NumberK41489
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer01/18/2023
03/09/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date02/15/2023
03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-