• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX430 PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX430 PATIENT MONITOR Back to Search Results
Model Number 866061
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted when additional information is received or upon completion of the investigation.
 
Event or Problem Description
The customer reported that their speaker was malfunctioning.The device was in clinical use when the issue was discovered but no patient harm was reported.
 
Event or Problem Description
This report is based on information provided by field service personnel and has been investigated by the philips complaint handling team.Philips has received a complaint about the intellivue mx430 patient monitor, indicating speaker malfunctioning.The following functional tests were performed: according to the fse, an analysis of the equipment was carried out and the failure was confirmed in the equipment.The following part need to be replaced: 1.00000,453564406331,iv2-flex main board.After replacing the part, the equipment no longer failed.Equipment released for use.Based on available evidence, the reported issue has been confirmed.Based on available information and the results of further analysis, no further action is required at this time.The device was operational after repairs were completed.The investigation concludes that no other action is needed at this time.If additional information is received, the claim file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX430 PATIENT MONITOR
Common Device Name
INTELLIVUE MX430 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen, MA 71034
GM   71034
6172455900
MDR Report Key15963718
Report Number9610816-2022-00602
Device Sequence Number14680620
Product Code MWI
UDI-Device Identifier00884838006379
UDI-Public00884838006379
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K090785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number866061
Device Catalogue Number866061
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer01/19/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-