STRYKER NEUROVASCULAR CORK TARGET 360 ULTRA 2MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035422060
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual/microscopic inspection confirmed that the main coil was intact.The coil delivery wire was seen to be kinked/bent.The coil delivery wire was seen to be broken/fractured.The coil introducer sheath was intact.Functional inspection was not required.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported main coil failed/unable to detach could not be confirmed; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was returned for analysis.The main coil was intact, the coil delivery wire was seen to be kinked/bent, the coil delivery wire was seen to be broken/fractured.The coil introducer sheath was intact.An assignable cause of procedural factors will be assigned to the reported main coil failed/unable to detach and as analyzed (aa) coil delivery wire broken/fractured during use, as these defects appear to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of handling damage will be assigned to aa code 'coil delivery wire kinked/bent', as this defect appears to be associated with handling of the product or portion of the product during the procedure.
 
Event or Problem Description
The subject device was returned for analysis and the device investigation revealed that the coil delivery wire was broken/fractured during use.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 ULTRA 2MM X 6CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15964288
Report Number3008881809-2022-00619
Device Sequence Number14687097
Product Code HCG
UDI-Device Identifier07613252239345
UDI-Public07613252239345
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/04/2024
Device Model NumberM0035422060
Device Catalogue NumberM0035422060
Device Lot Number22893086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2022
Initial Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient RaceAsian
-
-