• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM XS LFT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM XS LFT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6481-1-100
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
At the outpatient check-up on (b)(6) 2022, the x-rays showed rupture of the femoral stem with a disconnection between the mrh component of the femoral prosthetic shield and the femoral stem.On (b)(6) 2022, the complete revision surgery of the mrh size small femoral component and new 15x155mm femoral stem were performed.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.The following devices were also listed in this report: ti dur reg fluted stem 10x80mm; cat # 64786600; lot # m51s3.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a mrh femoral component was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "it is difficult to provide any real information regarding the case with the limited information.There is a hinge style implant in the knee.A single view unlabeled x-ray was provided (the second x-ray was obscured).On the x-ray the femoral stem appears angulated compared to the housing of the femoral implant.The event description indicated a revision was performed.Conclusion/assessment: apparent implant fracture or junctional fracture of the stem implant interface of a tumor style long stemmed hinged knee event confirmation: a fatigue failure of the stem at the housing interface of the femoral implant appears to have occurred.Root cause: a root cause for the implant failure cannot be determined from the materials provided." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fracture at the stem/femoral component interface.A review of the provided x-rays by a clinical consultant confirmed the event but was unable to identify root cause.It is unclear from the x-rays which component fractured, if not both.Further information such as return of the devices, additional pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
At the outpatient check-up on (b)(6) 2022, the x-rays showed rupture of the femoral stem with a disconnection between the mrh component of the femoral prosthetic shield and the femoral stem.On (b)(6) 2022, the complete revision surgery of the mrh size small femoral component and new 15x155mm femoral stem were performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRH KNEE FEM XS LFT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15964906
Report Number0002249697-2022-01806
Device Sequence Number12150121
Product Code KRO
UDI-Device Identifier07613327045109
UDI-Public07613327045109
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K994207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2015
Device Model Number6481-1-100
Device Catalogue Number64811100
Device Lot NumberSTM3L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer01/28/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age35 YR
Patient SexUnknown
-
-