Model Number 72203299
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during set up or inspection, when the spider was turned on, the arm fell directly to the ground.There was no delay and a backup device was available.There was no patient involvement.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.There was no relationship found between the device and the reported event.A visual inspection was performed on the product.No issues were noted.No accessories were included.No excess oil was noted on the outside of the device.A functional evaluation was performed.A test battery was utilized.No issues were noted.The device passed all functional testing.The device was left pressurized for approximately 3 hours.The device remained pressurized.100 movements and cycles were performed.No sudden de-pressurization events occurred.No issues with the clamping assembly were found.No issues with the main power switch or set up switch were found.A test foot switch jig and test switch drape jig were utilized.No issues were found.The piston migration was at 1.21 inches.No excess oil was noted within the enclosures.No problems were found.The reported complaint was not verified.The root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that could have contributed to the reported event include issues during setup of the device.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL, SPIDER 2, TENET 7615
Common Device Name
TABLE, OPERATING-ROOM, ELECTRICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15964991
Report Number1643264-2022-00471
Device Sequence Number4912692
Product Code GDC
UDI-Device Identifier00885554023534
UDI-Public00885554023534
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203299
Device Catalogue Number72203299S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-