| Model Number |
A2030-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified lesion in the right posterior tibial artery.A 3.0x120x150cm armada 14 balloon dilatation catheter (bdc) was advanced to the lesion for pre-dilatation.The balloon was inflated one time to 8 atmospheres (atm) when the balloon ruptured.There was no adverse patient effect and no clinically significant delay in the procedure.A 3.0x100x150cm armada 14 bdc was used as a replacement device.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and functional inspections were performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|