Model Number WA4KA400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to the olympus repair center.The repair inspection found a reflection on the image.Minor dents on the distal end edge and minor scratches on the gold plating at the distal area.However, the customer reported ¿broken during procedure¿ could not be confirmed.Olympus requested additional details regarding the reported event and is pending.The investigation is ongoing.However, if additional information becomes available, this report will be supplemented accordingly.In general, the end user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Event or Problem Description
Olympus was informed by the clinical engineering supervisor that the customer ultra rigid telescope was ¿broken during surgery¿.No death, injury or harm has been reported to olympus.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and because the phenomenon was not duplicated during device evaluation, the root cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE ¿ULTRA¿, 4 MM, 0°, WITH TROCAR TUBE CONNECTOR
Common Device Name
ULTRA TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15965448
Report Number9610773-2022-00631
Device Sequence Number5574408
Product Code HRX
UDI-Device Identifier04042761083003
UDI-Public04042761083003
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K160249
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA4KA400
Device Catalogue NumberWA4KA400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date12/12/2022
Supplement Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-