LIFESCAN EUROPE GMBH OT VERIO 2 METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Model Number 022-657
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Hypoglycemia (1912); Visual Impairment (2138); Ambulation or Postural Difficulties (2544)
Date of Event 12/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Similar complaints for this issue were trended including the reported meter.It was concluded that the number of complaints for the meter did not breach thresholds indicative of a systemic issue.In addition, similar complaints for this issue were trended for the test strip lot.It was concluded that the number of complaints for the lot did not breach thresholds indicative of a systemic issue.
 
Event or Problem Description
On (b)(6) 2022, the lay-user/patient contacted lifescan (lfs) france, alleging that their onetouch verio2 meter read inaccurately high compared to their feelings and/or normal readings.The complaint was classified based on the customer care agent (cca) documentation.The patient reported that the alleged meter inaccuracy began on (b)(6) 2022, at 6:30 pm when they obtained an alleged high result of ¿360 mg/dl¿ with the subject meter, and the issue continued for 1 week.The patient uses novomix 30 insulin (6-8 units morning and evening) to manage their diabetes.As a result of the alleged issue, the patient's nurse contacted their doctor who prescribed novorapid insulin to take in additional their normal medication.The patient changed their treatment as advised by their nurse and doctor, and on (b)(6) 2022, at 6:30 pm and (b)(6) 2022, at 5:30 am, they took 5 units of novorapid insulin in addition to 16 units of novomix 30 insulin.The patient indicated that on (b)(6) 2022, at 11:30 am, they developed symptoms of ¿couldn't see or walk properly¿ and ¿walked twisted¿.In response to the symptoms, the patient tested their blood glucose on another meter and obtained a result of ¿54 mg/dl¿.The patient self-treated with cake and 4 sugar cubes and felt better 10 minutes afterwards.At the time of troubleshooting, the cca confirmed the unit of measure was set correctly on the subject meter.The cca confirmed the test strip vial was intact and the test strips had been stored correct and were not expired or opened past their discard date.The cca noted the patient did not have control solution available to perform a quality control test.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly developed signs/symptoms suggestive of a serious injury adverse event after changing treatment, on the advice of their hcp, based on alleged inaccurate high results obtained with the subject meter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO 2 METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN SCOTLAND
beechwood park north
inverness IV2 3 ED
UK   IV2 3ED
Manufacturer Contact
simon palmer
beechwood park north
inverness IV2 3-ED
UK   IV2 3ED
1463383679
MDR Report Key15965653
Report Number3008382007-2022-04359
Device Sequence Number20221268
Product Code NBW
UDI-Device Identifier20353885008710
UDI-Public00353885008716
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K131363
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number022-657
Device Catalogue Number022-657
Device Lot Number5588093
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2022
Initial Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-