| Model Number |
1516-20-710 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Synovitis (2094); Joint Laxity (4526)
|
| Date of Event |
10/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for revision due to instability.Event is serious and is considered moderate.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2021.Date of revision: (b)(6) 2022.Date of event: (b)(6) 2022.(left knee).Treatment: revision; femoral component and insert were revised.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, the patient underwent a left knee revision for pain and instability.The patient had undergone physical therapy prior to the revision without improvement.Intraoperatively, chronic synovitis was identified.The femoral component was found to be well fixed.It was removed and downsized to a size 6.The revision was completed successfully.The femoral component and insert were the only implants revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint(b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: b5 and h6 (clinical codes).
|
| |
|
Search Alerts/Recalls
|