DEPUY ORTHOPAEDICS INC US ATTUNE CR FEM LT SZ 7 CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-00-107
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032)
Date of Event 08/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for revisions due to pain, stiffness, and adhesions.Event is serious and is considered moderate.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2020.Date of revision: (b)(6) 2022.Date of event: (b)(6) 2020.(left knee).Treatment: revision; femoral component and insert were revised.
 
Event or Problem Description
Operative notes received indicate that the patient underwent a revision of the femoral and tibial insert on (b)(6) 2022 for knee pain and stiffness.The surgeon states the patient has undergone extensive physical therapy and arthroscopic lysis of adhesions and manipulation to address severe stiffness secondary to ankylosis and arthrofibrosis.He continues to have a limited range of motion.Intraoperatively, extensive arthrofibrosis and dense scar tissue was identified.The femoral cruciate retaining component and cruciate retaining insert were exchanged for a femoral posterior stabilizing component and posterior stabilizing insert.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM LT SZ 7 CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15968645
Report Number1818910-2022-25192
Device Sequence Number5573568
Product Code JWH
UDI-Device Identifier10603295041085
UDI-Public10603295041085
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1504-00-107
Device Catalogue Number150400107
Device Lot Number9366236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer12/07/2022
01/18/2023
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date12/16/2022
01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 7 7MM; ATTUNE CR FEM LT SZ 7 CEM; ATTUNE FB TIB BASE SZ 6 CEM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE PS FB INSRT SZ 6 8MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-