• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE BFLEX 3.8 SINGLE-USE BRONCHOSCOPE; BRONCHOSCOPE (FLEXIBLE OR RIGID)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE BFLEX 3.8 SINGLE-USE BRONCHOSCOPE; BRONCHOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number 0570-0380
Medical Device Problem Code Erratic or Intermittent Display (1182)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported glidescope bflex 3.8 single-use bronchoscope was not returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history for the glidescope bflex 3.8 lot number "fu224100" identified one (1) similar complaint reported to verathon.Review of production records did not identify any anomalies that would cause or contribute to the reported image issue.Trending analysis for the glidescope bflex 3.8 single-use bronchoscopes does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous situation has been adequately captured and characterized by verathon.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event or Problem Description
A customer reported that during a patient procedure, using a glidescope bflex 3.8 single-use bronchoscope, the screen would completely freeze up when placing the bronchoscope in the et tube.They reported removing the bronchoscope to get the picture back, tried re-inserting it, and it froze again.The procedure was completed using another glidescope bflex 3.8 single-use bronchoscope which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE BFLEX 3.8 SINGLE-USE BRONCHOSCOPE
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key15970321
Report Number9615393-2022-00234
Device Sequence Number17378127
Product Code EOQ
UDI-Device Identifier00879123006394
UDI-Public010087912300639410FU22410017231007
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K193488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/07/2023
Device Model Number0570-0380
Device Catalogue Number0570-0396
Device Lot NumberFU224100
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Initial Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-