Model Number 1804275-18
Medical Device Problem Codes Material Rupture (1546); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat the ramus artery without calcification or tortuosity.The 2.75x18mm xience skypoint stent delivery system (sds) had no resistance during advancement and was inflated but was noted to not hold the pressure correctly and stopped inflating during the first inflation at an unknown pressure and either the balloon ruptured or the hypotube ruptured [kinked].The sds was removed, however, the half deployed stent was left in the left main artery.The stent was removed by using a snare device.The sds was prepared (air aspiration) outside the anatomy prior to use.Another same size device was used to complete the procedure.There was no adverse patient sequela and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported kinked shaft or material rupture; however, the reported treatment appears to be related to the operational context of the procedure.Additionally, it was reported that the partially deployed stent was removed from the anatomy by using a snare device.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15970354
Report Number2024168-2022-12566
Device Sequence Number5229355
Product Code NIQ
UDI-Device Identifier08717648233166
UDI-Public08717648233166
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number1804275-18
Device Catalogue Number1804275-18
Device Lot Number2032441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight113 KG
-
-