Model Number 530
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event of the back of the clip was bent and that bent part caused the hepatic artery to rupture and caused bleeding the hepatic artery is inconclusive.The device is not being returned and although photographic evidence was provided it shows no indication of the event.Therefore, a device malfunction cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised the following: the device is only for single use, and the ability to effectively clean and re-sterilize this single use device has not been established and subsequent re-use may adversely affect the performance, safety and/or sterility of the device.Additionally, the ifu goes on to state to always check to see that a clip is in the jaws before squeezing trigger.Closing of empty jaws on tissue may cause tissue damage.Always ensure trigger is squeezed completely and allowed to open to full initial position.Ensure the tissue or.Vessel fits completely within the confines of the clip or hemostasis may be compromised inspect the ligation site to ensure proper application and that hemostasis has been achieved.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The distributor reported on behalf of their customer that the 530, reflex elc,20m/l clps,10/11mm device, was being used during a robotic abdominal procedure on (b)(6) 2022 when it was reported, ¿in the process of connecting vessels of common hepatic artery with the reflex elc clip 10 mm during abdominal surgery using robot, the attached photo appeared and ruptured the common hepatic artery.It was sutured immediately but it was a very dangerous situation.During 6 hours of surgery, there was 200 cc of bleeding.¿ further assessment questioning found that ¿the back of the clip was bent and that bent part caused the hepatic artery to rupture and caused bleeding.¿ the procedure was completed, with an unknown amount of delay.The patient was hospitalized for an unknown amount of time and the patient¿s current status was reported as ¿patient is now healed.¿ it is unknown if any medical or surgical intervention was required for the patient.This report is being raised on the basis of injury due to patient¿s hepatic artery rupturing causing a loss of blood, which let to extended hospitalization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLEX ELC,20M/L CLPS,10/11MM
Common Device Name
APPLIER, HEMOSTATIC CLIP
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key15970776
Report Number1320894-2022-00240
Device Sequence Number4912873
Product Code HBT
UDI-Device Identifier30653405074810
UDI-Public(01)30653405074810(17)251101(10)20011021
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K924929
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/01/2025
Device Model Number530
Device Catalogue Number530
Device Lot Number202011021
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Initial Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-