• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN046313
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn#(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
It was reported that immediately after placement, 6 catheters on 6 different patients were non passable (occluded) and had to be replaced.The 6 events occurred in (b)(6) 2022, the exact date is unknown in each case.A patient was not harmed, each catheter was replaced immediately without an issue.
 
Event or Problem Description
It was reported that immediately after placement, 6 catheters on 6 different patients were non passable (occluded) and had to be replaced.The 6 events occurred in december 2022, the exact date is unknown in each case.A patient was not harmed, each catheter was replaced immediately without an issue.Associated mdrs: 3006425876-2022-01139, 3006425876-2022-01140, 3006425876-2022-01141, 3006425876-2022-01142, 3006425876-2022-01137.
 
Additional Manufacturer Narrative
(b)(4).The customer reported the catheter was occluded.The customer returned one epidural catheter.The returned sample was visually examined with and without magnification.Visual examination of the returned catheter revealed that the catheter appears used as biological material can be seen on the inner coils and adhesive can be seen on the outer extrusion.No other defects or anomalies were observed.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.10.The returned epidural catheter was inserted from the proximal end into a lab inventory snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 8.0ml/min, which is within the specification of 1ml/min minimum.No blockages were found.A device history record review was performed on the epidural catheter with no relevant findings.The reported complaint of the catheter being occluded could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.The returned sample passed a functional flow test and met flow rate specifications.There were no functional issues found with the returned sample.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194332672
MDR Report Key15971614
Report Number3006425876-2022-01138
Device Sequence Number6404632
Product Code BSO
UDI-Device Identifier00801902062512
UDI-Public00801902062512
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberIPN046313
Device Catalogue NumberJC-05400-DCS
Device Lot Number71F22K0259
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer01/24/2023
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexUnknown
-
-