Model Number NOVATION GXL LNR +5 LAT 40MM ID GROUP 4 CUPS
Medical Device Problem Codes Loss of or Failure to Bond (1068); Loss of Osseointegration (2408)
Health Effect - Clinical Codes Osteolysis (2377); Aseptic loosening (4883)
Date of Event 11/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported that a 69 yo male, initial right hip implanted on an unknown date, underwent a revision procedure on (b)(6) 2022, for reasons unknown.The surgeon removed the cup and liner and replaced with a competitor¿s devices.The patient was last known to be in stable condition following the event.There are no devices returning as they were sent to pathology.No further information.
 
Additional Manufacturer Narrative
After review of additional information received the following sections, have been updated accordingly: h6: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear/osteolysis: implanted with a lateralized liner and edge loading/impingement.A number of variables including use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential) could have all contributed.The most likely cause for the revision reported due to early prosthesis wear/osteolysis is a combination of the risk factors specified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Novation Acetabular Cup
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15971647
Report Number1038671-2022-01595
Device Sequence Number20397821
Product Code JDI
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNOVATION GXL LNR +5 LAT 40MM ID GROUP 4 CUPS
Device Catalogue Number136-40-54
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer03/23/2023
09/15/2026
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date03/28/2023
09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
62MM NOVATION CROWN CUP.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-