| Model Number |
HUM LNR IMPACT TIP 38M |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgical use, a 38mm impactor tip fractured during impaction of the liner.No parts or pieces fell into the patient wound site.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.The product is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on historical complaint investigations, the returned device, and the report event, the fractured humeral liner impactor tip reported was likely the result of a stress riser introduced during impaction, which led to crack initiation, propagation, and ultimate fracture of the device.Additionally, the device may have experienced material degradation if sterilized beyond the recommended parameters listed in the reprocessing instruction which could have led to the observed failure.D4: serial is unknown, unknown lot is 67739002.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury, nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|