| Model Number |
221283 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd bbl¿ trypticase¿ soy agar (soybean-casein digest agar) failed sterility test.The following information was provided by the initial reporter: customer states product failed sterility testing.Hazard, injury or erroneous results? no; hazard, injury or erroneous results details.
|
| |
|
Additional Manufacturer Narrative
|
|
The following corrections were made.Describe event or problem: it was reported that bd bbl¿ trypticase¿ soy agar (soybean-casein digest agar) plates were showed bacterial growth.No injuries reported.The following information was provided by the initial reporter: customer states product failed sterility testing.Visual inspection showed surface bacterial growth.
|
| |
|
Event or Problem Description
|
|
It was reported that bd bbl¿ trypticase¿ soy agar (soybean-casein digest agar) plates were showed bacterial growth.No injuries reported.The following information was provided by the initial reporter: customer states product failed sterility testing.Visual inspection showed surface bacterial growth.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6 investigation summary: during manufacturing of material 221283, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record review for batch 2277109 was satisfactory per internal procedures.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 2277109 for contamination.No retention samples for this batch were available for investigation.No photos or return samples were received for investigation of this complaint.Bd has identified a contamination trend for this product and the investigation found opportunities for bioburden reduction in the manufacturing process.Additional trainings are planned with an ongoing training review for cleaning processes.Improvement in observation of contamination is expected as the capa progresses.This complaint has been confirmed based on the trend.
|
| |
|
Event or Problem Description
|
|
It was reported that bd bbl¿ trypticase¿ soy agar (soybean-casein digest agar) plates were showed bacterial growth.No injuries reported.The following information was provided by the initial reporter: customer states product failed sterility testing.Visual inspection showed surface bacterial growth.
|
| |
|
Search Alerts/Recalls
|