| Model Number |
24700 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in moderately tortuous and severely calcified right superficial femoral artery.A 6.0mm x 100mm x 135cm-hybrid sterling balloon catheter was advanced for dilatation.However, during the first inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with a 6.0 x 200 charger balloon catheter.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 describe event or problem has been updated.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in moderately tortuous and severely calcified right superficial femoral artery.A 6.0mm x 100mm x 135cm-hybrid sterling balloon catheter was advanced for dilatation.However, during the first inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with a 6.0 x 200 charger balloon catheter.There were no patient complications reported.It was further reported that the balloon deflated and removed intact from the patient's body.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 describe event or problem has been updated.Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a longitudinal tear in the balloon 19mm from the tip and is approximately 50mm long.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a longitudinal tear in the balloon.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in moderately tortuous and severely calcified right superficial femoral artery.A 6.0mm x 100mm x 135cm-hybrid sterling balloon catheter was advanced for dilatation.However, during the first inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with a 6.0 x 200 charger balloon catheter.There were no patient complications reported.It was further reported that the balloon deflated and removed intact from the patient's body.
|
| |
|