• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IDEAL IMPLANT INCORPORATED IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IDEAL IMPLANT INCORPORATED IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT Back to Search Results
Model Number 24001
Medical Device Problem Codes Material Perforation (2205); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Alleged deflation.
 
Event or Problem Description
Deflation.
 
Additional Manufacturer Narrative
Macroscopic and microscopic examination of the explanted device revealed a minute particle of cured silicone within the shell causing outer shell failure and deflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDEAL IMPLANT STRUCTURED BREAST IMPLANT
Common Device Name
SALINE-FILLED BREAST IMPLANT
Manufacturer (Section D)
IDEAL IMPLANT INCORPORATED
14881 quorum drive
suite 925
dallas TX 75254
Manufacturer (Section G)
VESTA INC.,
9900 south 57th street
franklin WI 53132
Manufacturer Contact
robert hamas
14881 quorum drive
suite 925
dallas, TX 75254
MDR Report Key15973655
Report Number3011491947-2022-00348
Device Sequence Number5692991
Product Code FWM
UDI-Device Identifier10851795006022
UDI-Public(01)10851795006022(241)24001(17)221209
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/09/2022
Device Model Number24001
Device Catalogue Number240
Device Lot Number709106
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date04/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-