On (b)(6) 2022, the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprosthesis and gore® excluder® aaa endoprosthesis.After all stent grafts were implanted, intra-operative imaging reportedly revealed a proximal type i endoleak.Touch-up ballooning was repeated, however, the endoleak was not resolved.An aortic extender endoprosthesis (gore® excluder® conformable aaa endoprosthesis) was implanted proximally to treat the type ia endoleak.According to the report, the aortic extender endoprosthesis partially and unintentionally covered the left renal artery.As the blood flow of the left renal artery was not delayed, it was decided no additional treatment was required.The endoleak was resolved, and the procedure was concluded without any further reported complications.
|
|
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device remains implanted and, therefore, was not available for direct analysis by gore.It should be noted that, per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, improper endoprosthesis component placement, and occlusion of device or native vessel.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|