• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number CXA280005
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 11/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprosthesis and gore® excluder® aaa endoprosthesis.After all stent grafts were implanted, intra-operative imaging reportedly revealed a proximal type i endoleak.Touch-up ballooning was repeated, however, the endoleak was not resolved.An aortic extender endoprosthesis (gore® excluder® conformable aaa endoprosthesis) was implanted proximally to treat the type ia endoleak.According to the report, the aortic extender endoprosthesis partially and unintentionally covered the left renal artery.As the blood flow of the left renal artery was not delayed, it was decided no additional treatment was required.The endoleak was resolved, and the procedure was concluded without any further reported complications.
 
Additional Manufacturer Narrative
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device remains implanted and, therefore, was not available for direct analysis by gore.It should be noted that, per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, improper endoprosthesis component placement, and occlusion of device or native vessel.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
damon jackson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15973774
Report Number3007284313-2022-02264
Device Sequence Number6894331
Product Code MIH
UDI-Device Identifier00733132650958
UDI-Public00733132650958
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/11/2024
Device Model NumberCXA280005
Device Catalogue NumberCXA280005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Initial Report FDA Received Date12/13/2022
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexMale
-
-