| Model Number |
FX4508SWC |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Osseointegration (2646)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The doctor reported the implant was removed due to failure to osseointegrate approximately 1 month after implant placement.Bone quality around the area was type iv, and oral hygiene around the implant was good.The doctor reported bone resorption was observed during the implant removal surgery.The patient will need another surgical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Please see attached summary memo.
|
| |
|
Search Alerts/Recalls
|