TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number 1002717 |
Device Problem
False Alarm (1013)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an altitude alarm occurred while the customer was not outside of labeled operating altitude range.Customer's blood glucose was within range (value not provided).Reportedly, the customer cleaned the speaker holes and cleared the alarm.
|
|
Search Alerts/Recalls
|
|
|