The product was returned and radiographs were provided and both confirmed the complaint.The patient's postoperative physical activity is unknown and it is unknown if the patient suffered a fall.No torque handle information was provided and it is unknown if hand tightening was utilized.Examination found the right connector arm separated from the housing assembly and deformed as well as fractured in half.Rod lock screws were separated from the assembly with extreme rod contact damage on the inferior side.Though no definitive root cause can not be determined the damage observed and reported time in-situ suggest excessive loading related to postoperative physical activity or incomplete final lock down resulting in postoperative loosening, toggle and subsequent fracture.A follow up report will be submitted should additional information be received.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant components." "warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.The vuepoint oct system and vuepoint ii oct system can also be linked to 3.5mm to 6.25mm rods of posterior pedicle screw and rod systems via the rod to rod connectors or transition rods.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of the implants.Metallic internal fixation devices cannot withstand the activity levels and/or loads equal to those placed on normal, healthy bone.These devices are not designed to withstand the unsupported stress of full weight or load bearing alone.Care should be taken to insure that all components are ideally fixated prior to closure." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed"."pre-operative warnings: care should be used during surgical procedures to prevent damage to the devices and injury to the patient"."post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.".
|