• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-610
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 11/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event or Problem Description
On (b)(6) 2022, senseonics was made aware of an adverse event where user experienced false hypoglycemia due to inaccuracies in sensor readings.
 
Additional Manufacturer Narrative
Based on the investigation analysis, the system may have displayed temporary mismatch between the sensor reading and fingerstick measurement on the 9th day after insertion.The insertion site was still recovering after insertion, and as the sensor stabilized in the body, the system started to show better agreement between the sensor readings and fingerstick measurements.The system is currently in use, and the sensor data shows that it is performing within expectations.Per case notes, user was not symptomatic.User didn't require any medical treatment and didn't need to self resolve.Per dms, user is currently using the system with up to date information.No further resolution was found necessary for this complaint.D4.Updated additional device information.H3.Device evaluated by manufacturer?: yes.H4.Device manufacturer date updated to 01jun2022.H6.Investigation findings updated to 114.H6.Investigation conclusions updated to 4315.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key15974693
Report Number3009862700-2022-00185
Device Sequence Number14879386
Product Code QHJ
UDI-Device Identifier00817491022929
UDI-Public817491022929
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/01/2023
Device Model Number102208-610
Device Catalogue NumberFG-5901-01-001
Device Lot Number129895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient SexMale
-
-