• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY FLOWGATE2 8F X 95CM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY FLOWGATE2 8F X 95CM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 90495
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the procedure, the subject balloon ruptured inside the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No further information is available.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.H3 device evaluated by mfg ¿ updated.H3 summary attached - updated.D4 expiration date - added.D10 product available to stryker ¿ updated.D10 returned to manufacturer on ¿ updated there are controls in the manufacturing process to ensure the product met specifications upon release.During visual/microscopic inspection, the balloon catheter was found to be kinked/bent.The balloon was found to be damaged/ruptured.During functional inspection, balloon burst/ruptured functional test ¿ balloon inflation was not attempted due to condition of the device.Balloon damage was confirmed during visual inspection.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event was confirmed during analysis.The device failed to meet specification when received, due to the damage noted.It was reported that during the last intervention the balloon ruptured.Additional information states that the device was prepared for use as per the directions for use, the device confirmed to be in good condition during preparation/prior to use on the patient, continuous flush was set up and maintained throughout the clinical procedure and the patient¿s anatomy was moderately tortuous.The device was returned, and the balloon was noted to be damaged/ruptured, confirming the reported event, the balloon catheter shaft was also kinked.It is probable that the device was damaged during advancement through the patient¿s anatomy causing the reported event.An assignable cause of procedural factors will be assigned to the as reported and as analyzed ¿balloon burst/ruptured during use¿ and to the as analyzed ¿ balloon catheter kinked/bent¿, as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.
 
Event or Problem Description
It was reported that during the procedure, the subject balloon ruptured inside the patient's anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWGATE2 8F X 95CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15975104
Report Number3012931345-2022-00223
Device Sequence Number20563520
Product Code DQY
UDI-Device Identifier00815742004953
UDI-Public00815742004953
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K153729
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2024
Device Model Number90495
Device Catalogue Number90495
Device Lot Number0000125096
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2022
Supplement Date Received by Manufacturer02/24/2023
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date03/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-