| Model Number |
48-30-00 |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Livanova deutschland manufactures the s5 system.The incident occurred in united states.A livanova field service engeneet was dispatched to the customer.The motor ends stage board (90-305-760), motor control hms (90-305-770) were replaced.Device was returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland received a report that a s5 console presented error fault in motor controller (429) in pump 2 during a procedure.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The review of similar complaints on the affected device has been performed, showing that no other complaints have been registered on roller pumps in the last 24 months.The most probable root cause was an electrical/electronic deviation of the replaced components.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|