The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: only the distal fragment of the nail was returned.The nail had broken in the webs of the oblong drill hole at distal.The breakage surface showed evidences of a fatigue fracture like lines of rest and missing plastic deformation.With the help of a microscope a secondary fracture could be identified close to the breakage edge of the web.Scuffed off coating resp.Non scuffed off coating inside the drill holes for the locking screw identified not all drill holes had been occupied.Otherwise no deficiencies determined.Dimensional inspection indicates that the device is within the specifications.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance with the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.In the case presented a 66-year-old female patient, no other data available, had been treated with above implant on (b)(6) 2021 due to not named reasons.On (b)(6) 2022, 18 months post initial treatment, the implant was revised due to breakage.Medical records were and / or op-reports were not made available.With given information it could not be determined if the bone had healed, it could not be determined if the patient had been compliant.The question of weight bearing in general could not be cleared, either.The breakage pattern on the item returned showed a fatigue fracture indicated by lines of rest, missing plastic deformation and a secondary crack.The general appearance of the breakage surfaces lead to to the conclusion that implant had been highly loaded.If any additional information is provided, the investigation will be reassessed.
|