Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.The mitch trh acetabular system components are oem devices.Stryker orthopaedics is not the legal manufacturer of the mitch devices.However, since this device was used in conjunction with a stryker device they are being listed as associated devices: std mitch trh cp sz 50/56, cat #: mac-9988-5056, lot #: fm059145; mitch trh md hd sz 50-4, cat #: mmh-9988-0450, lot #: fm067775.
|
|
Reported event: an event regarding wear, corrosion, and abnormal ion level involving an exeter stem was reported.Wear and corrosion were confirmed via evaluation of the returned device.Abnormal ion level was not confirmed.Method & results: product evaluation and results: visual inspection: visual inspection of the returned device was performed as part of the material analysis: "on visual examination, material loss and areas of discoloration were observed around the intersection of the neck and trunnion of the stem." material analysis: a material analysis was performed which concluded the following: "review of the exeter v40 stem, catalogue # 0580-1-441, lot code g1660672 was completed using visual examination, stereo microscopy examination and energy dispersive spectroscopy examination.Material loss was observed at the intersection between the neck of the stem and the trunnion, consistent with in-vivo wear, potentially related to the stem trunnion losing taper lock with the head taper.Elemental analysis confirmed the presence of oxygen in the areas with material loss, indicating corrosion.No manufacturing related defects were observed on the device surfaces examined." clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to elevated metal ion levels.A material analysis was performed on the returned device and it concluded the following: "review of the exeter v40 stem, catalogue # 0580-1-441, lot code g1660672 was completed using visual examination, stereo microscopy examination and energy dispersive spectroscopy examination.Material loss was observed at the intersection between the neck of the stem and the trunnion, consistent with in-vivo wear, potentially related to the stem trunnion losing taper lock with the head taper.Elemental analysis confirmed the presence of oxygen in the areas with material loss, indicating corrosion.No manufacturing related defects were observed on the device surfaces examined." the elevated metal ion levels could not be confirmed as no medical records were provided for review.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|