MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00539270
Medical Device Problem Codes Break (1069); Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a flexima biliary stent was used during an endoscopic retrograde cholangiopancreatography procedure in the biliary drainage, performed on (b)(6) 2022.During the procedure, when the physician attempted to deploy the stent, the guide catheter broke and the stent failed to deploy.Another flexima biliary stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Block h6: medical device problem code a0401 captures the reportable event of guide catheter break.The returned flexima biliary stent was analyzed, and a visual evaluation noted that the device looks in good condition overall and no issues were found.The stent was still loaded.The functional inspection was performed, and the stent deploy successfully.The reported event was not confirmed.Based on the provided and collected information, the device met all manufacturing requirements, and no abnormalities were reported during the manufacturing process.After product analysis of the returned device, it was determined that the device was free of obvious issues and during functional inspection it worked as intended, the unit returned did not show evidence of either the alleged issue or any defect that could have contributed to the event reported.Therefore, no problem detected will be the overall conclusion code for the complaint.
 
Event or Problem Description
It was reported to boston scientific corporation that a flexima biliary stent was used during an endoscopic retrograde cholangiopancreatography procedure in the biliary drainage, performed on (b)(6) 2022.During the procedure, when the physician attempted to deploy the stent, the guide catheter broke and the stent failed to deploy.Another flexima biliary stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15975999
Report Number3005099803-2022-07179
Device Sequence Number6959822
Product Code FGE
UDI-Device Identifier08714729162582
UDI-Public08714729162582
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K965147
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2024
Device Model NumberM00539270
Device Catalogue Number3927
Device Lot Number0028181279
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer12/26/2022
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date01/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
Patient Weight66 KG
-
-