Model Number 24704
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device entrapment occurred.The patient presented with tibial artery blockage and underwent balloon angioplasty.After gaining vascular access via the femoral artery, a 3.0mm x 80mm x 150cm coyote balloon catheter was advanced for dilatation.However, during introduction, the balloon got stuck on a 300cm angled tip v-14 control wire guidewire and would not move further into the target lesion.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a coyote balloon catheter with guidewire stuck inside of it.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed buckling to the shaft 4cm and 6.2cm from the tip.There is also buckling under the strain relief.Microscopic examination revealed no additional damages.The guidewire was pulled lightly but it does not appear to be moving so the guidewire was left inside the device.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found damage to the device that may have contributed to the reported event.
 
Event or Problem Description
It was reported that device entrapment occurred.The patient presented with tibial artery blockage and underwent balloon angioplasty.After gaining vascular access via the femoral artery, a 3.0mm x 80mm x 150cm coyote balloon catheter was advanced for dilatation.However, during introduction, the balloon got stuck on a 300cm angled tip v-14 control wire guidewire and would not move further into the target lesion.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15976093
Report Number2124215-2022-49500
Device Sequence Number5784377
Product Code DQY
UDI-Device Identifier08714729796541
UDI-Public08714729796541
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2025
Device Model Number24704
Device Catalogue Number24704
Device Lot Number0028843147
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date02/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
Patient Weight68 KG
-
-